Background and objectives: According to the manufacturer, the Bispectral Index (BIS) has a processing time delay of 5-10s. Studies addressing this have suggested longer delays. We evaluated the time delay in the Bispectral Index response.
Methods: Based on clinical data from 45 patients, using the difference between the predicted and the real BIS, calculated during a fixed 3minutes period after the moment the Bispectral Index dropped below 80 during the induction of general anesthesia with propofol and remifentanil.
Results: The difference between the predicted and the real BIS was in average 30.09±18.73s.
Conclusion: Our results may be another indication that the delay in BIS processing may be much longer than stated by the manufacture, a fact with clinical implications.
Description
Keywords
Monitor do índice bispectral Bispectral index monitor Profundidade da anestesia Depth of anesthesia Consciência Awareness Tempo de atraso Time delay
Citation
Ferreira AL, Mendes JG, Nunes CS, Amorim P. Avaliação do tempo de atraso do índice bispectral na resposta à indução da anestesia: estudo observacional [Evaluation of Bispectral Index time delay in response to anesthesia induction: an observational study]. Rev Bras Anestesiol. 2019;69(4):377-382. doi:10.1016/j.bjan.2019.03.008
Publisher
Elsevier